Primer probe: Master Mix lot variation

Each Pathogen detection assay was evaluated for detection consistency with

each of the probes in the multiplex mixture (

stx1/stx2; eae,

and O157).

Figure 2.

mericon

E. coliSTEC O-type

Figure 2: Primer-probe and master mix lot variation for

mericon

E. coliSTEC

O-typeassay

Ct values for all primer/probe (PP) and Master Mix (MM) lot combinations for

the

mericon

E. coli STEC O-type Pathogen Detection Assaywere not

significantly different, indicating that there were no effects of individual lots of

primer/probes and Master Mix on the detection results. The assay is a

multiplex giving results in the green channel for the non-O157 Big 6 STEC O-

0 5 10 15 20 25 30

E.coli

O157:H7

E.coli

O157

E.coli

O26:H-

E.coli

O111:H-

E.coli

O145 Citrobacter

freundii

Ct

Bacterial Strain

mericon®

E. coli STEC O-type

: O-type

PP1/MM1

PP2/MM1

PP3/MM1

PP1/MM2

PP2/MM2

PP3/MM2

PP1/MM3

PP2/MM3

0.00

5.00

10.00

15.00

20.00

25.00

30.00

E.coli

O157:H7

E.coli O157 E. coli

O111:H-

E.coli O145 Citrobacter

freundii

Ct

Bacterial Strain

mericon®

E. coli STEC O-type: O157:H7

PP1/MM1

PP2/MM1

PP3/MM1

PP1/MM2

PP2/MM2

PP3/MM2

PP1/MM3

PP2/MM3

OMAMAN-36 E/ AOAC PTM Report

ERP Use Only

January 2017

AOAC Research Institute