34

of

76

For treatment combination 9, there were 3 presumptive positives out 10 replicates

for the variations evaluated. For the 5 uninoculated test portions, there were 0

presumptive positives out of 5 replicates.

For the low inoculation level, a POD value of 0.30 was obtained with a 95%

confidence interval of (0.11, 0.60). All 5 uninoculated test portions were negative

with a POD value of 0.00 with a 95% confidence interval of (0.00, 0.43).

Detailed results for the robustness tests are presented in Tables 21and 22ofthe

Appendix.

Product Consistency (Lot to Lot) and Stability Evaluation

Results Lot to lot consistency study

Primer probe: Master Mix lot variation

mericonE. coli O157 Screen Plus Pathogen Detection Assay

Each Pathogen detection assay was evaluated for detection consistency with each of the

probes in the multiplex mixture (

stx1/stx2; eae,

and O157

Figure1.

mericon

E. coliO157 Screen Plus

0

5

10

15

20

25

30

35

40

E.coli

O157:H7

E.coli O157 E.coli

O26:H-

E.coli

O111:H-

Citrobacter

freundii

Ct

Bacterial Strain

mericon

E. coli O157 Screen Plus: stx1/stx2

PP1/MM1

PP2/MM1

PP3/MM1

PP1/MM2

PP2/MM2

PP3/MM2

PP1/MM3

PP2/MM3

PP3/MM3

OMAMAN-36 E/ AOAC PTM Report

ERP Use Only

January 2017

AOAC Research Institute

Expert Review Panel Use Only