35

of

76

Figure 1: Primer-probe and master mix lot variation for

mericon

E. coliO157

Screen Plus assay.

Ct values for all primer/probe (PP) and Master Mix (MM) lot combinations for

the

mericon

E. coliO157Screen Plus Pathogen Detection Assaywere not

significantly different, indicating that there were no effects of individual lots of

primer/probes and Master Mix on the detection results. The assay is a

multiplex giving results in the green channel for

stx1/stx2,

in the orange

channel for

O157¸

and in the crimson channel for

eae.

Results for

Citrobacter freundii

(a closely related but non-

E. coli

bacterial strain)are

shown.

Lot to Lot Study:

mericonE. coli STEC O-Type Pathogen Detection Assay

0

5

10

15

20

25

30

35

40

E.coli

O157:H7

E.coli

O157

E.coli

O26:H-

E.coli

O111:H- Citrobacter

freundii

Ct

Bacterial Strain

mericon

E. coli O157 Screen Plus

:

O157

PP1/MM1

PP2/MM1

PP3/MM1

PP1/MM2

PP2/MM2

PP3/MM2

PP1/MM3

PP2/MM3

PP3/MM3

0

5

10

15

20

25

30

35

40

E.coli

O157:H7

E.coli O157 E.coli

O26:H-

E.coli

O111:H-

Citrobacter

freundii

Ct

Bacterial Strain

mericon

E. coli O157 Screen Plus: eae

PP1/MM1

PP2/MM1

PP3/MM1

PP1/MM2

PP2/MM2

PP3/MM2

PP1/MM3

PP2/MM3

PP3/MM3

OMAMAN-36 E/ AOAC PTM Report

ERP Use Only

January 2017

AOAC Research Institute