Table of Contents Table of Contents
Previous Page  401 / 453 Next Page
Information
Show Menu
Previous Page 401 / 453 Next Page
Page Background

Page

39

of

76

Figure4. Cycler variation for

mericon

E. coliSTEC O-type assay.

Ct values for all Cycler combinations using consistent primer probe:master

mix lots for the

mericon

E. coliSTEC O-type Pathogen Detection Assaywere

not significantly different, indicating that there were no effects of different

cyclers on the detection results.

Consistency Study: Bacteria Plus Prep lot variation

Each Pathogen detection assay was evaluated for detection consistency with

each of the probes in the multiplex mixture (

stx1/stx2; eae, and O157)

using 3

different lots of the

mericon

DNA Bacteria Plus kits.

Figure 5

; mericon

E. coliO157 Screen Plus Prep lot variation

0.00

5.00

10.00

15.00

20.00

25.00

30.00

E.coli

O157:H7

E.coli O157 E.coli

O126:H2

E.coli O145 Citrobacter

freundii

Ct

Bacterial Strain

mericon®

E. coli STEC O-type

: O157:H7

Cycler

RG1

RG2

RG3

0.00 5.00 10.00 15.00 20.00 25.00 30.00

E.coli

O157:H7

E.coli

O157

E.coli

O111:H-

E.coli

O126:H2

E.coli

O145 Citrobacter

freundii

Ct

Bacterial strain

mericon

E. coli O157

Screen Plus: stx1/stx2

Sample prep

prep 1

prep 2

prep 3

OMAMAN-36 E/ AOAC PTM Report

ERP Use Only

January 2017

AOAC Research Institute

Expert Review Panel Use Only