Dual Lab Validation of a Method for the Determination of Fructans in Infant 

Formula & Adult Nutritionals /        

25 Jul 2016

39 

Table 3 

Fructan content (% m/m) in the applied fructan standards (duplicate 

averages) measured with the AOAC 997.08 and with the submitted 

SPIFAN methods. 

Fructan standard 

ingredient 

AOAC 997.08 

method 

Submitted SPIFAN method 

duplicate

average 

duplicates 

average 

Oraftie HP 

91.7    

93.0

92.7 

94.6    90.5 

92.6 

Oraftie P95 

85.8    

85.4 

85.6 

84.7    80.8 

82.8 

Nutraflora P95 

86.0    

86.4 

86.2 

88.6    85.6 

87.1 

The duplicate averages as measured with the AOAC 997.08 method are used in the standard addition 

experiments for calculating the exact amount of added pure fructan. 

b.

Calibration fit 

In this method quadratic calibration models are used for quantifying the glucose and the fructose 

contents. Glucose is the minor constituent  in fructans. Therefore the working range for glucose is at 

lower levels than for fructose. In the standard daily routine just 5 calibration samples for glucose and 

fructose are applied, ranging for glucose from 0.05– 20 mg/l and for fructose from 0.90 – 90 mg/l.  

For testing the calibration models 12 calibration samples are used. Two times with the ISC5000_E 

instrument and one time with the ICS3000_D instrument the calibration models were established for 

this validation study. The results are summarized in table 4.  