Dual Lab Validation of a Method for the Determination of Fructans in Infant 

Formula & Adult Nutritionals /        

25 Jul 2016

42 

Table 5 Summarizing results of the repeatability and reproducibility 

investigations in the reconstituted samples. 

sample 

no 

sample 

description 

number 

of 

analyses 

average

content

% m/m 

SD

% m/m 

RSD

r

meet 

target 

SD

R

% m/m 

RSD

R

meet 

target 

Child formula powder 

5 x 2 + 1 x 1 

0,287 

0,004 

1,4 

yes 

0,021 

7,4 

yes 

Toddler formula 

powder milk based 

6 x 2 

0,223 

0,003 

1,6 

yes 

0,013 

6,3 

yes 

10 

infant formula powder 

milk based 

6 x 2 

0,263 

0,003 

1,1 

yes 

0,017 

6,5 

yes 

12 

child formula powder 

6 x 2 

0,282 

0,003 

1,1 

yes 

0,016 

6,3 

yes 

14 

infant formula powder 

FOS/GOS based 

6 x 2 

0,036 

0,004 

8,9 

no 

0,005 

13,8 

no 

19 

Adult nutritional RTF 

high fat 

6 x 2 

0,507 

0,013 

2,6 

yes 

0,021 

4,6 

yes 

As can been seen in the experimental data in Annex C, there was an outlier value measured in the 

duplicate at day 1 in sample 1. This outlier was detected with Grubb’s outlier test. In the summarizing 

statistical evaluation in table 5, this outlying value was rejected. Therefore the number analyses in 

sample 1 is just 5 x 2 + 1 x1 instead of 6 x 2. This means that the SD

r

 and RSD

r

are based on 5 

duplicates instead of 6 duplicates. And the SD

R

 and RSD

R

 are based on 11 values instead of 12 

measured fructan concentrations. 

The RSD

r

 and RSD

R

of sample no 14 “infant formula powder FOS/GOS based” don’t meet the target 

values of respectively of 6 and 12%. However this is not surprisingly because the fructan content in 

this sample is very low, in the range of the LOQ of the method. And for that reason the RSD

r

and RSD

R

are somewhat higher than the target values in the SMPR 2014.02. 

e.

Accuracy 

The same 6 FOS containing samples (after reconstition) in the SPIFAN sample kit as selected for the 

precision study have been applied for the accuracy study. The average fructan content in these 

samples was already determined in the precision study. Those date are used in the accuracy study as 

the fructan content before spiking the samples with a pure fructan ingredient. 

The established purity of the fructan ingredients Orafti HP, Orafti P95 and Nutraflora P95, applied for 

the standard additions, are given already in table 3. 

The addition solutions are prepared by dissolving  about 1.1 gram (weighted with an accuracy of 1 mg) 

in 100 ml water. For all samples except sample 14, the same amounts of fructan have been added in 

the addition experiments. The fructan content in sample 14 (infant formula powder FOS/GOS based) 

was very low, near the LOQ value. Therefore in case of this sample a much lower addition level was 

applied. 

For the samples 1, 9, 10, 12 and 19 the addition at low level a volume of 0.70 ml was added to 4.00 ml 

of the reconstituted sample and for addition at the high level a volume of 2.00 ml was added to 4.00 

ml of the reconstituted sample. For sample 14 the added volume of the standard fructan solutions was 