Parameter

Acceptance Criteria

R2

≥ 0.99

For ≤ 10 μg/l : 70-125 %

10-1000 μg/l : 75-120 %

Accuracy

70 - 125 %

Precision (RSD %)

≤ 15 %

Least/Most intesne

ion ratio

Maximum tolerance (relative)

0.50 - 1.00

± 30 %

0.20 - 0.50

± 30 %

0.10 - 0.20

± 30 %

<0.10

± 30 %

RT ratio

0.95 - 1.05

Concentration

Recovery limits, %

≤ 10 μg/kg

70 - 125

1.0 mg/kg

75 - 120

10 mg/kg

80 - 115

0.01 g/100g

85 - 110

0.1 g/100g

90 - 108

1.0 g/100g

92 - 105

10 g/100g

95 - 102

100 g/100g

98 - 101

Concentration

Repeatability (RSDr), %

≤ 10 μg/kg

15

1.0 mg/kg

8

10 mg/kg

6

0.01 g/100g

4

0.1 g/100g

3

1.0 g/100g

2

10 g/100g

1.5

100 g/100g

1

Concentration

Repeatability (RSD

R

), %

≤ 10 μg/kg

32

1.0 mg/kg

16

10 mg/kg

11

0.01 g/100g

8

0.1 g/100g

3

1.0 g/100g

2

10 g/100g

1.5

100 g/100g

1

Page 4

Accuracy

Table : Validation parameters and Acceptance Criteria for B- Vitamins

Validation Parameter

Criteria

Repeatability Precision (RSDr)

Reproducibility Precision (RSD

R

)

Linearity/Calibration fit

Back calculated

values accuracy

LOQ

Specificity

Ion Ratio

BVit-03 (February 2016)

FOR ERP USE ONLY