Table of Contents Table of Contents
Previous Page  167 / 670 Next Page
Information
Show Menu
Previous Page 167 / 670 Next Page
Page Background

Validation

Parameter

Test Group Name

Date

Test Name

Data Path File

Matrix

Raw-Data File Ref.

Limit

CRM Data File Ref.

Equipment

Analyst

Test Method

Operator

Repeatability at LOQ 5 ug/l standard

Sample Blank No.

Mean Retention Time (min)

m/z

Signal (ug/l)

1

1.823

265.2>122.2

4.608

Mean (μg/l)

4.44

2

1.823

265.2>122.2

4.785

SD

0.3339

3

1.823

265.2>122.2

4.453

LOD (Mean +3 sigma)

5.45

4

1.823

265.2>122.2

4.814

LOQ (Mean + 10sigma)

7.78

5

1.823

265.2>122.2

4.194

RSD %

7.51

6

1.823

265.2>122.2

4.135

Standard Conc. (μg/l)

5

7

1.823

265.2>122.2

4.675

Accuracy %

88.9

8

1.823

265.2>122.2

4.774

9

1.823

265.2>122.2

4.099

Accuracy = 70 - 125 %

10

1.823

265.2>122.2

3.911

RSD% = <15%

Page 21

Criteria

Caluclation

TUV/BLR/F-360

2015\January\Vitamins\01.01.2015

01.01.2015

AOAC- SPIFAN-2012-VITAMINS

Limit of Quantification

B-Complex vitamins

Thiamine

Aqueous

LC-MS/MS 6460A (BLR/PS-CRS/006.03)

Prasad BKVR/Kesav Kumar N

Manjunath K/Prasad BKVR

TUV/BLR/F-99

BVit-03 (February 2016)

FOR ERP USE ONLY

DO NOT DISTRIBUTE