Table of Contents Table of Contents
Previous Page  169 / 670 Next Page
Information
Show Menu
Previous Page 169 / 670 Next Page
Page Background

Validation

Parameter

Test Group Name

Date

Test Name

Data Path File

Matrix

Raw-Data File Ref.

Limit

CRM Data File Ref.

Equipment

Analyst

Test Method

Operator

Repeatability at LOQ 5 ug/l standard

Sample Blank No.

Mean Retention Time (min)

m/z

Signal (ug/l)

1

2.315

123.2>80.3

4.949

Mean (μg/l)

4.73

2

2.315

123.2>80.3

4.916

SD

0.1886

3

2.315

123.2>80.3

4.539

LOD (Mean +3 sigma)

5.29

4

2.315

123.2>80.3

4.639

LOQ (Mean + 10sigma)

6.61

5

2.315

123.2>80.3

4.896

RSD %

3.99

6

2.315

123.2>80.3

4.580

Standard Conc. (μg/l)

5

7

2.315

123.2>80.3

4.966

Accuracy %

94.5

8

2.315

123.2>80.3

4.491

9

2.315

123.2>80.3

4.725

Accuracy = 70 - 125 %

10

2.315

123.2>80.3

4.555

RSD% = <15%

Page 23

Criteria

Caluclation

AOAC- SPIFAN-2012-VITAMINS

Limit of Quantification

B-Complex vitamins

Niacinamide

Aqueous

LC-MS/MS 6460A (BLR/PS-CRS/006.03)

01.01.2015

2015\January\Vitamins\01.01.2015

Prasad BKVR/Kesav Kumar N

TUV/BLR/F-360

TUV/BLR/F-99

Manjunath K/Prasad BKVR

BVit-03 (February 2016)

FOR ERP USE ONLY

DO NOT DISTRIBUTE